Calibrated UV sensor with communication via RS485 (Modbus/ZCON) as plug-in sensor for UV systems according to DVGW W 294 / DIN 19294.
Processing via PLC or ZED UV Monitor in “W/m²”, “mW/cm²” or in “%”
Recalibration is carried out by ZED,
optional: “in-field recalibration” for OEM customers for on-site recalibration using ZED SmartMeter without an external service provider
Features
- available for low-pressure and medium-pressure applications
- measuring range for low pressure: 2...500W/m²
- measuring range for medium pressure: 20...3000W/m²
- precise detection of UVC irradiance over the entire specified measuring range - other measuring ranges can also be realized on request
- Multi-sensoring - parallel operation of the sensors on the common RS485 bus
- for use with a measuring window in accordance with DVGW 294-2 / DIN 19294 (e.g. ZED MF001)
Technical Specification
supply voltage: | 12...24V DC |
---|---|
sensitive element: | silicon carbide diode (SiC) |
spectral range: | 210 - 290nm |
operation temperatur: | 0 - 60°C (32 - 140°F) |
max. pressure / IP code: | not waterproof without measurement window MF001 IP64 on terminal ends if connected to appropriate sensor cable |
body material: | stainless steel 1.4305 |
mounting: | use with MF001 (MF001: pipe thread ISO228 G1) |
wiring: | ready-to-use ZED sensor cables available in several lengths |
Display Units & Accessories
- ZED UV Monitors
- UV Monitor PRO11
- UV Monitor PRO16
- UV Monitor PRO30D
- Control units
- ZCONmini II
- via RS485/ModBus RTU: PLC or PC
- Accessories
- Measurement Window MF001
- Sensor PC-Software
- ZED Sensor Configurator
- Radiometer
- SmartMeter
- TinyMeter